Video-assisted thoracoscopic surgery in device closure of atrial septal defect


Department of Cardiovascular Surgery, Union Hospital, Fujian Medical University, Fuzhou, 350001, P. R, China

Date of Web Publication20-Sep-2013

Correspondence Address:
Hua Cao
Department of Cardiovascular Surgery, Union Hospital, Fujian Medical University, Fuzhou, 350001, P. R
China
Login to access the Email id


DOI: 10.4103/1817-1737.118498

PMID: 24250740

Rights and Permissions



How to cite this article:
Cao H, Chen Q, Zhang GC, Hu YN, Xu F. Video-assisted thoracoscopic surgery in device closure of atrial septal defect. Ann Thorac Med 2013;8:234-5

How to cite this URL:
Cao H, Chen Q, Zhang GC, Hu YN, Xu F. Video-assisted thoracoscopic surgery in device closure of atrial septal defect. Ann Thorac Med [serial online] 2013 [cited 2020 Aug 14];8:234-5. Available from: http://www.thoracicmedicine.org/text.asp?2013/8/4/234/118498


Sir,

Atrial septal defect (ASD) is a common congenital heart disease. Compared with surgical and percutaneous closure method [1],[2] recently along with the development of thoracoscopic technique in cardiac surgery, we have combined the thoracoscopic technique and the transthoracic minimally invasive closure technique and in order to achieve a more esthetically acceptable outcome. In this study, we reported totally thoracoscopic surgical closure of ASD, to treat ten cases of ASDs from January 2012 to September 2012.

Ten patients were included in the study, six males and four females, with a mean age of 27.2 ± 10.6 years (age range 16-42 years). All patients were diagnosed with an ostium secundum ASD without any other intracardiac deformities. The diameter of the defect ranged from 16 mm to 24 mm. The present study was approved by the Ethics Committee of our university and was conducted according to the tenets of the Declaration of Helsinki. The ASD occluder was made from Lifetech Scientific (Shenzhen) Co., Ltd., China.

In our method, the patient was placed supine with the right side of the chest elevated about 45΀. A parasternal incision 1 cm in length was made in the right anterior chest wall. The thoracic cavity was reached through the 4 th intercostal space. One-lung ventilation was employed during surgery. Another incision 1 cm in length was made at the intersection of the midaxillary line and the 4 th intercostal space for the thoracoscope [Figure 1]. The pericardium was opened with an electric hook and a traction suture was placed on the pericardium. The suture line was pulled out through the incision on the midaxillary line to expose the right atrium. A purse-string suture was applied in the anterolateral wall of the right atrium. Systemic heparin (1 mg/kg) was used. A fine needle was aspirated from the center of the purse. The guidewire and delivery sheath were placed in succession [Figure 2]. Under ultrasound guidance, the delivery sheath was extended to the left atrium through the ASD. An occluder was slowly advanced via the sheath. The left atrial umbrella was open within the left atrium and pulled back tightly and then the right atrial umbrella was open. Ultrasound examination was carried out to confirm that the umbrella closure was tight, there was no residual shunt and the structures and functions of the superior vena cava, inferior vena cava, pulmonary vein, coronary sinus opening, mitral valve and tricuspid valve were intact. Finally, the occluder and the delivery sheath were removed in turn and the purse-string suture in the right atrium was ligated.
Figure 1: The surgical field

Click here to view
Figure 2: The sheath positioned from the right atrial free wall by thoracoscope

Click here to view


The closure procedure was performed successfully in all 10 patients. The operative time ranged from 60 min to 90 min (mean 70.5 ± 10.6 min). The time for intracardiac closure ranged from 5 min to 10 min (mean 7.1 ± 1.5 min). The size of the occluder for the ASD ranged from 20 mm to 28 mm (mean 24.4 ± 3.2 mm). Perioperative arrhythmia, including supraventricular tachycardia and atrial flutter, occurred in two patients. Cardiac rhythm returned to normal in a relatively short period of time after medication was administered. In those who had a successful attempt, the overall immediate complete closure rate was 100%.

Similar to some reports on transcatheter or surgical ASD closure, [1],[2] we also achieved high technical success and good therapeutic outcomes. Our method has several advantages. It was easy to learn and cost-acceptable in the Third World nations. It can completely avoid the application of cardiopulmonary bypass and its small incision (1 cm) is cosmetically preferable to patients. It can be used for patients with relatively large defects and thus, extends the indication of defect closure. It can completely avoid X-ray exposure. It is most important to note that in case of failed closure, our method can be quickly converted to cardiopulmonary bypass surgery in the same operating room without transporting the patient from the interventional suite to the operating room.

We believe that transthoracic minimally invasive ASD closure is safe and effective, as is demonstrated by the experience we have documented. [3],[4] On this basis, have we introduced the thoracoscopic technique and obtained a perfect view inside the right chest cavity by left-sided one-lung ventilation. [5],[6] Through the 1-cm incision in the anterior chest wall, we can incise the pericardium, carry out purse-string suture and place the closure sheath. In addition, the approach used for thoracoscopy also can be applied to insert the drainage tube, further reducing the extent of the incision. The most difficult aspects of this technique are the manipulation of instruments under thoracoscopy and the technique of atrial septal closure using the occluder. The operative time in the current study was relatively long compared with that of conventional transthoracic minimally invasive closure. The long operative time results mainly from the inexperience of those performing the manipulation under thoracoscopy during surgery. With increased experience in the surgical procedure, the operative time is expected to be shortened. The magnification effect of thoracoscopy allows us to observe bleeding in the surgical field more clearly and assists with hemostasis during surgery. In the current study, no large pleural or pericardial effusion was observed. We suggest that the operator should have sufficient experience in heart surgery and should be familiar with intraoperative ultrasound imaging. Compared with percutaneous closure of an ASD, our procedure can be converted to open surgery immediately, if necessary, which makes the surgeon more confident.

In our opinion, we believe that this procedure cannot replace traditional surgery and percutaneous interventional procedures; however, it could be an option for the treatment of ASD. Currently, various surgical procedures are available to most patients. Therefore, the selection of surgical procedure should be based on individual considerations such as the condition and the needs of the patient, the size of ASD, the experience of the physician and the availability of equipment in the hospital.

In summary, we believe that totally thoracoscopic surgical closure of ASD is a safe and effective procedure, that it is minimally invasive and that it can achieve an esthetically acceptable outcome.

 
   References Top

1.Kirklin JW, Barratt-Boyes BG. Atrial septal defect and partial anomalous pulmonary venous connection. In: Cardiac Surgery. 2 nd ed. Churchill Livingstone, New York; 1993. p. 609-44.  Back to cited text no. 1
    
2.Yew G, Wilson NJ. Transcatheter atrial septal defect closure with the Amplatzer septal occluder: Five-year follow-up. Catheter Cardiovasc Interv 2005;64:193-6.  Back to cited text no. 2
    
3.Chen Q, Chen LW, Cao H, Zhang GC, Chen DZ, Zhang H. Intraoperative device closure of atrial septal defects with inferior vena cava rim deficiency: A safe alternative to surgical repair. J Thorac Cardiovasc Surg 2011;141:631-6.  Back to cited text no. 3
    
4.Chen Q, Cao H, Zhang GC, Chen LW, Chen DZ. Safety and feasibility of intra-operative device closure of atrial septal defect with transthoracic minimal invasion. Eur J Cardiothorac Surg 2012;41:121-5.  Back to cited text no. 4
    
5.Yu SQ, Cai ZJ, Cheng YG, Duan DW, Xu XZ, Chen WS, et al. Video-assisted thoracoscopic surgery for congenital heart disease. Asian Cardiovasc Thorac Ann 2002;10:228-30.  Back to cited text no. 5
    
6.Chang CH, Lin PJ, Chu JJ, Liu HP, Tsai FC, Lin FC, et al. Video-assisted cardiac surgery in closure of atrial septal defect. Ann Thorac Surg 1996;62:697-701  Back to cited text no. 6
    


    Figures

  [Figure 1], [Figure 2]



 

Top
Print this article  Email this article
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Article in PDF (461 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    References
    Article Figures

 Article Access Statistics
    Viewed1650    
    Printed41    
    Emailed0    
    PDF Downloaded273    
    Comments [Add]    

Recommend this journal