ORIGINAL ARTICLE
Year : 2011  |  Volume : 6  |  Issue : 4  |  Page : 217-220

Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure


1 Department of Respiratory Diseases, Hospital of Sestri Levante, Chiavarese, Italy
2 Department of Public Health, ASL4 Chiavarese, Italy

Correspondence Address:
Antonello Nicolini
Department of Respiratory Diseases, Hospital of Sestri, Levante
Italy
Login to access the Email id


DOI: 10.4103/1817-1737.84776

PMID: 21977067

Rights and Permissions

Background: Non-invasive measurement of oxygenation is a routine procedure in clinical practice, but transcutaneous monitoring of PCO 2 (PtCO 2 ) is used much less than expected. Methods : The aim of our study was to analyze the value of a commercially available combined SpO 2 /PtCO 2 monitor (TOSCA-Linde Medical System, Basel, Switzerland) in adult non-invasive ventilated patients with acute respiratory failure. Eighty critically ill adult patients, requiring arterial blood sample gas analyses, underwent SpO 2 and PtCO 2 measurements (10 min after the probe was attached to an earlobe) simultaneously with arterial blood sampling. The level of agreement between PaCO2 - PtCO 2 and SaO 2 - SpO 2 was assessed by Bland-Altman analyses. Results : Both, SaO 2 from blood gas analysis and SpO 2 from the transcutaneous monitor, and PaCO 2 and PtCO 2 were equally useful. No measurements were outside of the acceptable clinical range of agreement of ± 7.5 mmHg. Conclusions : The accuracy of estimation of the TOSCA transcutaneous electrode (compared with the "gold standard" blood sample gas analysis) was generally good. Moreover, TOSCA presents the advantage of the possibility of continuous non-invasive measurement. The level of agreement of the two methods of measurement allows us to state that the TOSCA sensor is useful in routine monitoring of adults admitted to an intermediate respiratory unit and undergoing non-invasive ventilation.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed6604    
    Printed166    
    Emailed2    
    PDF Downloaded568    
    Comments [Add]    
    Cited by others 11    

Recommend this journal