ORIGINAL ARTICLE
Year : 2011  |  Volume : 6  |  Issue : 1  |  Page : 33-37
Comparison and agreement between venous and arterial gas analysis in cardiopulmonary patients in Kashmir valley of the Indian subcontinent


1 Department of Internal and Pulmonary Medicine, Sher-i-Kashmir Institute of Medical Sciences, Srinagar, Kashmir, India
2 Department of Clinical Biochemistry, Sher-i-Kashmir Institute of Medical Sciences, Srinagar, Kashmir, India

Date of Submission16-Sep-2010
Date of Acceptance16-Oct-2010
Date of Web Publication28-Dec-2010

Correspondence Address:
Parvaiz A Koul
Department of Internal and Pulmonary Medicine, Sher-i-Kashmir Institute of Medical Sciences, Soura, Srinagar - 190 011, Kashmir
India
Login to access the Email id


DOI: 10.4103/1817-1737.74274

PMID: 21264169

Rights and Permissions

   Abstract 

Background: Arterial blood gas (ABG) analysis is routinely performed for sick patients but is fraught with complications, is painful, and is technically demanding.
Objective: To ascertain agreement between the arterial and peripheral venous measurement of pH, pCO 2 , pO 2 , and bicarbonate levels in sick patients with cardiopulmonary disorders in the valley of Kashmir in the Indian subcontinent, so as to use venous gas analysis instead of arterial for assessment of patients.
Setting: Sher-i-Kashmir Institute of Medical Sciences, Srinagar, Kashmir, a 650-bedded tertiary care hospital in North India located at an altitude of 1584 m.
Methods: One hundred patients who required ABG analysis were admitted. Peripheral venous blood was drawn within 5 min of an ABG measurement, and the samples analyzed immediately on a point of care automated ABG analyzer. Finger pulse oximetry was used to obtain oxygen (SpO 2 ) saturation. Data were analyzed using Pearson correlation and bias (Bland Altman) methods.
Results: The venous measurements of pH, pCO 2 , pO 2 and bicarbonate, and the digital oxygen saturation were highly correlated with their corresponding arterial measurements. Bland Altman plots demonstrated a high degree of agreement between the two corresponding sets of measurements with clinically acceptable differences. The difference in pO 2 measurements was, however, higher (-22.34 ± 15.23) although the arterial saturation and finger oximetry revealed a good degree of agreement with clinically acceptable bias.
Conclusion: Peripheral venous blood gas assessment in conjunction with finger pulse oximetry can obviate the routine use of arterial puncture in patients requiring ABG analysis.


Keywords: Arterial, blood gases, COPD, correlation, pH, venous


How to cite this article:
Koul PA, Khan UH, Wani AA, Eachkoti R, Jan RA, Shah S, Masoodi Z, Qadri SM, Ahmad M, Ahmad A. Comparison and agreement between venous and arterial gas analysis in cardiopulmonary patients in Kashmir valley of the Indian subcontinent. Ann Thorac Med 2011;6:33-7

How to cite this URL:
Koul PA, Khan UH, Wani AA, Eachkoti R, Jan RA, Shah S, Masoodi Z, Qadri SM, Ahmad M, Ahmad A. Comparison and agreement between venous and arterial gas analysis in cardiopulmonary patients in Kashmir valley of the Indian subcontinent. Ann Thorac Med [serial online] 2011 [cited 2019 Jun 26];6:33-7. Available from: http://www.thoracicmedicine.org/text.asp?2011/6/1/33/74274


Acid base analysis is an essential tool in acute care medicine, yielding valuable information about a diverse group of disorders. Arterial blood gas (ABG) is the gold standard method for acquiring this information. Although used to evaluate many respiratory and metabolic conditions, ABG analysis is not without drawbacks. Blood sample for ABG is obtained by arterial puncture. The risks associated with this procedure include pain, hematoma, aneurysm formation, thrombosis or embolization, needle stick injuries to health care providers, and sometimes even reflex sympathetic dystrophy. [1],[2],[3],[4] The procedure is technically more demanding and is more painful. Over the years researchers have searched for alternatives to ABG sampling. Recently, studies have shown that values of acid-base balance measured in central or peripheral venous blood correlate well with those measured in arterial blood, at least for values of pH, bicarbonate, and carbon dioxide tension. [5],[6],[7],[8] Venous blood is easier to obtain, the procedure being less painful, and there is an advantage of blood being drawn at a time when the patient is getting sampled for other investigations. While there have been few reports from different regions of the world, no study has specifically sampled cardiopulmonary patients from our part of world situated at a moderately high altitude of 1584 m where lung disorders form one of the commonest causes of hospital admissions, especially during the winter months. [9] As a result there is a huge requirement of performing ABG analysis. An alternative to the arterial puncture would be quite helpful in this setting, and the current study is aimed to determine the extent of correlation of arterial and venous blood gas (VBG) analysis with a view to identifying whether the venous samples can be used as an alternative to arterial values in the clinical management of patients presenting with various cardiopulmonary disorders.


   Methods Top


This prospective study was conducted in the Department of Internal and Pulmonary Medicine of Sher-i-Kashmir Institute of Medical Sciences, a 650-bedded tertiary care university hospital facility in the Northern Indian State of Jammu and Kashmir. A total of 100 randomly selected patients, who were adjudged to require ABG analysis by the treating physician, were enrolled. The patients included 63 cases admitted with chronic obstructive pulmonary disease (COPD) and exacerbation, 20 cases with community acquired pneumonia/sepsis, 11 cases with congestive heart failure, 5 cases with interstitial lung disease, 3 cases with post-tubercular fibrosis, and 1 patient with obstructive sleep apnea. All patients had arterial blood sampled which was drawn via an arterial puncture into a heparinized syringe. Simultaneously, venous blood in another heparinized syringe was also sampled via antecubital vein without application of any tourniquet. Venous and arterial blood samples were taken within 5-min of each other. In addition, the measurement of oxygen saturation (SpO 2 ) was obtained from a finger pulse oximeter (Nonin pulse oximeter). Blood was analyzed immediately for values of acid-base, oxygenation and electrolyte status. The ABG and VBG analyzes were run on an immunoradiometric assay point-of-care blood gas analyzer that is operated by the Department of Biochemistry (Synthesis IL-45 analyzer, Instrumentation Laboratory, Lexington, MA). The assay is routinely used for analyzing blood gases in patients, and its accuracy is validated daily.

All the documented data were analyzed using Med Cal Statistical Software (version 10.4). The mean and 95% confidence intervals were calculated for each arterial and venous variable and for difference between them. The strength of the relationship between the arterial and venous gas values was assessed with the Pearson product-moment correlation coefficient test. The Bland-Altman limits of agreement were determined by plotting the difference between two paired values (arterial and venous) against their mean, thus creating a bias plot. The 95% limits of agreement represents the mean difference between each pair of venous and arterial values ±1.96 SD, and they estimate by how much a venous value is likely to differ from the criterion standard, the arterial value. This is the accepted method for assessing the agreement between two tests and represents a clinically relevant measure of comparison. [10],[11] A P value of <0.05 for means and Pearson r was considered significant. Informed consent for participation in the study was obtained from all the participants, and the study was approved by the Institute Postgraduate and Ethics Committee.


   Results Top


A total of 100 patients who required ABG analysis were enrolled in this study, and 200 sets of blood gas values were analyzed. The mean age of the subjects was 44 ± 19.63 and the patients included 63 males and 37 females. [Table 1] depicts the mean, SD, and range of the values of arterial pH, venous pH, arterial paCO 2 , venous pCO 2 , arterial and venous bicarbonate, and arterial and venous pO 2 . In addition, the oxygen saturation as determined by ABG analysis (SaO 2 ) and by digital pulse oximetry (SpO 2 ) is depicted in [Table 1]. Arterial pH and venous pH were found to be correlated significantly (Pearson correlation coefficient r = 0.88, 95% confidence limits of r = 0.82-0.92, P < 0.001). There was a strong correlation between the arterial and venous pCO 2 (r = 0.92, 95% confidence limits of r = 0.88-0.94, P < 0.0001), bicarbonate values (r = 0.32, 95% confidence limits 0.13-0.49, P = 0.0012), and pO 2 levels (r = 0.45, 95% confidence limits of r = 0.27-0.59, P < 0.0001). A significant correlation was also observed between SaO 2 and SpO 2 (r = 0.59, 95% confidence limits of r = 0.45-0.71, P < 0.001).
Table 1 :Summary statistics showing mean, standard deviation (SD), 95% confidence limits of mean (CI), median, and range of the various studied parameters


Click here to view


Bland Altman bias plots of the average of the arterial and venous measurements, and the difference between them are depicted in [Figure 1],[Figure 2],[Figure 3],[Figure 4],[Figure 5]. The bias plot for each of the variables showed excellent agreement with 95% limits of agreement being in acceptably narrow range in case of all the parameters. The mean bias in pH was -0.030 (SD = 0.030, 95% limits of agreement = -0.088 to 0.028); bias in pCO 2 was -4.05 (SD = 4.48, 95% limits of agreement = -12.8 to 4.7), and in bicarbonate levels 086 (SD = 2.15, 95% limits of agreement = -3.35 to 5.09). Although the arterial and venous pO 2 were significantly correlated with each other (r = 0.45, P < 0.0001), the bias between the two was -22.34 (SD = 15.23, 95% limits of agreement = -52.2 to 7.5). However, the bias between measured oxygen saturation (SaO 2 ) and digital pulse oximetry (SpO 2 ) was low (-1.44, SD = 8.82, 95% limits of agreement = -18.7 to 15.8), these results thus demonstrate an acceptably low bias with excellent agreement between the arterial and peripheral venous values.
Figure 1 :Bland Altman plots of arterial and venous pH (Average versus Difference)

Click here to view
Figure 2 :Bland Altman plots of arterial (pCO2 A) and venous pCO2 (pCO2 V) (Average vs. Difference)

Click here to view
Figure 3 :Bland Altman plots of arterial and venous HCO3 values (Average vs. Difference)

Click here to view
Figure 4 :Bland Altman plots of arterial and venous pO2 (Average vs. Difference)

Click here to view
Figure 5 :Bland Altman plots of measured oxygen saturation (SaO2) and on pulse oximetry (SpO2) (Average vs. Difference)

Click here to view



   Discussion Top


Our data have demonstrated a high level of correlation between the arterial and venous measurements of pH, pCO 2 , and bicarbonate level with a high degree of agreement and clinically acceptable difference on Bland Altman difference plotting. Although there is a significant correlation between the arterial and venous measurements of pO 2 level, the difference of -22.34 ± 15.23 is not clinically acceptable enough to support uniform usage of venous pO 2 instead of arterial measurements in clinical situations. However, SaO 2 and SpO 2 were highly correlated and demonstrated excellent agreement and a clinically acceptable difference (-1.44 ± 8.82). Caution is required for using correlation coefficients to describe the agreement between two quantitative variables. A high correlation is definitely required but does not necessarily imply agreement, for which the line of agreement should be a 45º (x = y) line. Bland Altman plots are used instead to describe the agreement between two quantitative variables, the difference of the paired two observations plotted against the mean of two measurements and 95% of the data points recommended to lie within ±2D of the mean difference. [12]

Several researchers have looked at the possibility of using the venous or capillary data instead of ABGs and have reported differing results. Gambino compared "arterialized" capillary blood with arterial samples from the brachial artery in patients undergoing pulmonary function tests and found no difference in pH and CO 2 content. [13] Subsequently Zahn and Weil [14] examined the relationship between the arterial and central venous pH and concluded that central venous pH was a reliable indicator of arterial pH. However , Adrogue et al. [15] emphasized that both arterial and central venous sampling are needed in patients with critical hemodynamic compromise. Other investigators have also supported the use of venous measurements instead of arterial measurements for pH and gas analysis [16],[17] while others [18] have resisted the idea. Arterial or venous data have been rarely noted to influence the physician's treatment in emergency in case of diabetic ketoacidosis, [16],[19] and venous values have been demonstrated to have excellent agreement with arterial values in cases of uremic acidosis. [20]

All the samples of the patients in our study were taken in the supine position. In a study comparing the effect of various body positions on the ABGs, there were essentially no significant differences in pO 2 among supine, sitting, and standing positions in asymptomatic patients of different age-groups or in patients with COAD [21] even as another study demonstrated that because of unpredictable postural changes in PaO 2 in patients with COPD, body position should be noted for ABG measurements and should be kept constant for valid comparison of serial measurements. [22]

The majority of our patients had lung disorders with COPD forming the commonest condition. Earlier COPD patients on mechanical ventilation [23] and those managed without mechanical ventilation [12] have been studied and both the groups of investigators reported that VBG measurements could be used instead of ABG analysis for assessment of the patients. The patients included in our study resided at a moderately high altitude. It is known that the barometric pressure decreases with distance above the Earth's surface in an approximately exponential manner. The pressure at 5500 is only about one-half the normal 760 mmHg, so the pO 2 of moist inspired gas is (380-47) + 0.2093 = 70 mmHg (partial pressure of water vapor equals 47 mmHg). In spite of hypoxia associated with high altitude, some 15 million people live at elevations over 3050 m depicting a remarkable degree of acclimatization at these altitudes. The permanent residents who live at these altitudes have a degree of hyperventilation that compensates for the degree of hypoxia, but at the expense of decreased PCO 2 . The hyperventilation results from the hypoxic stimulation of the peripheral chemoreceptors. Graham and Houston [24] have reported that patients with COPD (without CO 2 retention) tolerate high altitudes of up to 2000 m well, complaining of fatigue only. [24],[25] This explains the behavior of our patients at the moderate altitudes of Kashmir. Whether altitude of this magnitude plays a role in the blood gases needs to be studied in greater detail subsequently. However in the context of the current study, even if does, it would presumably affect the arterial and venous gases similarly.

We used peripheral venous blood in our study for VBG assessment. While analysis of central venous blood has been previously used for data comparison of VBG versus ABG [5],[15],[26] sampled in central venous catheters in ICU patients, the sampling of peripheral blood allows greater flexibility in obtaining data in patients without central venous access. Our results fortify the sparse data [5],[7],[27] regarding the view that peripheral VBG analysis can be used for fairly accurate assessment of arterial pH, pCO 2 , and bicarbonate values with clinically acceptable differences between the arterial and the venous values. Although there is a high correlation between the arterial and venous pO 2 values, the difference is large for total clinical agreement and thus suggesting that one cannot be swapped for the other without loss of important detail. However, SaO 2 and SpO 2 have high degree of correlation and high degree of agreement with clinically acceptable difference. Thus if venous pH is combined with routine finger pulse oximetry, the ABG analysis might be minimized to a great degree in cardiopulmonary patients. Additional studies may be necessary to address the feasibility and reproducibility of such testing.

 
   References Top

1.McGillivray D, Ducharme FM, Charron Y, Mattimoe C, Treherne S. Clinical decision making based on venous versus capillary blood gas values in the well perfused child. Ann Emerg Med 1999;34:58-63.  Back to cited text no. 1
[PUBMED]    
2.Mortensen J. Clinical sequelae from arterial needle puncture, cannulation, and incision. Circulation 1967;35:1118-23.  Back to cited text no. 2
    
3.Kelly AM, Kyle E, McAlpine R. Venous pCO 2 and pH can be used to screen for significant hypercarbia in emergency patients with acute respiratory disease. J Emerg Med 2002;22:15-9.  Back to cited text no. 3
[PUBMED]  [FULLTEXT]  
4.Criscuolo C, Nipper G, Buchalter S. Reflex sympathetic dystrophy following arterial blood gas sampling in the Intensive Care setting. Chest 1995;108:578-80.  Back to cited text no. 4
    
5.Toftegaard M, Rees SE, Andreassen S. Correlation between acid-base parameters measured in arterial blood and venous blood sampled peripherally, from vena cavae superior, and from the pulmonary artery. Eur J Emerg Med 2008;15:86-91.  Back to cited text no. 5
[PUBMED]  [FULLTEXT]  
6.Malatesha G, Singh NK, Bharija A, Rehani B, Goel A. Comparison of arterial and venous pH, bicarbonate, pCO 2 and pO 2 in initial emergency department assessment. Emerg Med J 2007;24:569-71.  Back to cited text no. 6
[PUBMED]  [FULLTEXT]  
7.Kelly AM, McAlpine R, Kyle E. Venous pH can safely replace arterial pH in the initial evaluation of patients in the emergency department. Emerg Med J 2001;18:340-2.  Back to cited text no. 7
[PUBMED]  [FULLTEXT]  
8.Rees SE, Hansen A, Toftegaard M, Pedersen J, Kristiensen SR, Harving H. Converting venous acid-base and oxygen status to arterial in patients with lung disease. Eur Respir J 2009;33:1141-7.  Back to cited text no. 8
[PUBMED]  [FULLTEXT]  
9.Koul PA, Wahid A, Jan RA. Long term oxygen therapy in COPD: Review of the guidelines and a critical audit of usage in the Kashmir Valley of the Indian Subcontinent. Kashmir Med J 2008;2:112-8.  Back to cited text no. 9
    
10.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-10.  Back to cited text no. 10
[PUBMED]    
11.Bland JM, Altman DG. Comparing methods of measurement: Why plotting difference against standard method is misleading. Lancet 1995;346:1085-7.  Back to cited text no. 11
[PUBMED]  [FULLTEXT]  
12.Chan YH. Biostatistics: Correlation analysis. Singapore Med J 2003;44:59-64.  Back to cited text no. 12
    
13.Gambino SR. Collection of capillary blood for simultaneous estimation of arterial pH, CO 2 content, pCO 2 and oxygen saturation. Am J Clin Pathol 1961;35:175-83.   Back to cited text no. 13
    
14.Zahn RL, Weil MH. Central venous blood for monitoring pH and pCO 2 in the critically ill patient. J Thorac Cardiovasc Surg 1966;52:105-11.  Back to cited text no. 14
[PUBMED]    
15.Adrogue HJ, Rashad MN, Gorin AB, Yacoup J, Madias NE. Assessing acid-base in circulatory failure. Differences in arterial and central venous blood. N Engl J Med 1989;320:1312-6.  Back to cited text no. 15
    
16.Brandenburg MA, Dire DJ. Comparison of arterial and venous blood gas values in the initial emergency department evaluation of patients with diabetic ketoacidosis. Ann Emerg Med 1998;4:459-65.  Back to cited text no. 16
    
17.Kelly AM, McAlpine R, Kyle E. Agreement between bicarbonate measured on arterial and venous blood gases. Emerg Med Autralas 2004;16:407-9.  Back to cited text no. 17
    
18.Gennis PR, Skovron ML, Aronson ST, Gallagher HJ. The usefulness of peripheral venous blood in estimating acid base status in acutely ill patients. Ann Emerg Med 1985;14:845-9.  Back to cited text no. 18
    
19.Ma OJ, Rush MD, Godfrey MM, Gaddis G. Arterial blood gas results rarely influence emergency physician management of patients with suspected diabetic ketoacidosis. Acad Emerg Med 2003;10:836-41.  Back to cited text no. 19
[PUBMED]  [FULLTEXT]  
20.Gokel Y, Paydas S, Koseoglu Z, Alparslan N, Seydaoglu G. Comparison of blood gas and acid-base measurements in arterial and venous blood samples in patients with uremic acidosis and diabetic ketoacidosis in the emergency room. Am J Nephrol 2000;20:319-23.  Back to cited text no. 20
[PUBMED]  [FULLTEXT]  
21.Marti C, Ulmer WT. Absence of effect of the body position on arterial blood gases. Respiration 1982;43:41-4.  Back to cited text no. 21
[PUBMED]    
22.Ries AL, Kaplan RM, Chang G. Effect of posture on arterial oxygenation in patients with chronic obstructive pulmonary disease. Respiration 1992;59:317-21.  Back to cited text no. 22
    
23.Chu YC, Chen CZ, Lee CH, Chen CW, Chang HY, Hsiue TR. Prediction of arterial blood gas values from venous blood gas values in patients with acute respiratory failure receiving mechanical ventilation. J Formos Med Assoc 2003;102:539-43.   Back to cited text no. 23
[PUBMED]    
24.Graham WG, Houston CS. Short term adaptation to moderate altitude: Patients with chronic obstructive lung disease. JAMA 1978;240:1491-4.  Back to cited text no. 24
[PUBMED]    
25.Karrer W, Schmid T, Wutrich O, Baldi W, Gall E, Portman HR. Respiration in patients with chronic lung disease at 500 and 1500 meters above sea level. Schweiz Med Wochenschr 1990;120:1584-9.  Back to cited text no. 25
    
26.Middleton P, Kelly AM, Brown J, Robertson M. Agreement between arterial and central venous values for pH, bicarbonate, base excess, and lactate. Emerg Med J 2006;23:622-4.  Back to cited text no. 26
[PUBMED]  [FULLTEXT]  
27.Ak A, Ogun CO, Bayir A, Kayis SA, Koylu R. Prediction of arterial blood gas values from venous blood gas values in patients with acute exacerbation of chronic obstructive pulmonary disease. Tohoku J Exp Med 2006;210:285-90.  Back to cited text no. 27
[PUBMED]  [FULLTEXT]  


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1]

This article has been cited by
1 Peripheral venous and arterial blood gas analysis in adults: are they comparable? A systematic review and meta-analysis
Anthony L Byrne,Michael Bennett,Robindro Chatterji,Rebecca Symons,Nathan L Pace,Paul S Thomas
Respirology. 2014; : n/a
[Pubmed] | [DOI]
2 Base Excess as a Predictor for Injury Severity in Pediatric Trauma Patients
Yochai Levi,Igor Jeroukhimov,Kobi Peleg,Michael Rozenfeld,Itai Shavit,Eran Kozer
The Journal of Emergency Medicine. 2013; 45(4): 496
[Pubmed] | [DOI]
3 Agreement between arterial and venous blood gases in emergency medical care: A systematic review
Kelly, A.-M.
Hong Kong Journal of Emergency Medicine. 2013; 20(3): 166-171
[Pubmed]
4 Comparison of Accuracy: Bland-Altman analysis of alternative examinations in the field of respiratory medicine
Umeda, A. and Okada, Y.
Respiration and Circulation. 2012; 60(8): 840-848
[Pubmed]
5 Are arterial and venous samples clinically equivalent for the estimation of pH, serum bicarbonate and potassium concentration in critically ill patients?
Herrington, W.G., Nye, H.J., Hammersley, M.S., Watkinson, P.J.
Diabetic Medicine. 2012; 29(1): 32-35
[Pubmed]
6 Are arterial and venous samples clinically equivalent for the estimation of pH, serum bicarbonate and potassium concentration in critically ill patients?
W. G. Herrington,H. J. Nye,M. S. Hammersley,P. J. Watkinson
Diabetic Medicine. 2012; 29(1): 32
[Pubmed] | [DOI]
7 Agreement and differences between venous and arterial gas analysis
Chhabra, S.K.
Annals of Thoracic Medicine. 2011; 6(3): 154
[Pubmed]



 

Top
Print this article  Email this article
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (579 KB)
    Citation Manager
    Access Statistics
    Reader Comments